Home  |   Archive  |   Online Submission  |   News & Events  |   Subscribe  |   APFA  |   Society  |   Contact Us  |   中文版
Search   
 
Journal

Ahead of print
Authors' Accepted
    Manuscripts
new!
Current Issue
Archive
Acknowledgments
Special Issues
Browse by Category

Manuscript Submission

Online Submission
Online Review
Instruction for Authors
Instruction for Reviewers
English Corner new!

About AJA

About AJA
Editorial Board
Contact Us
News

Resources & Services

Advertisement
Subscription
Email alert
Proceedings
Reprints

Download area

Copyright licence
EndNote style file
Manuscript word template
Guidance for AJA figures
    preparation (in English)

Guidance for AJA figures
    preparation (in Chinese)

Proof-reading for the
    authors

AJA Club (in English)
AJA Club (in Chinese)

 
Abstract

Volume 13, Issue 4 (July 2011) 13, 636–639; 10.1038/aja.2011.19

In vitro comparison of the vaporesection of human benign prostatic hyperplasia using 70- and 120-W 2-μm lasers

Guang-Heng Luo1,2, Shu-Jie Xia1 and Zhao-Lin Sun2

1 Department of Urology, Shanghai First People's Hospital Affiliated to Shanghai Jiaotong University, Shanghai 200080, China
2 Guizhou Province People's Hospital, Guiyang 550002, China

Correspondence: Dr SJ Xia, (xsjurologist@163.com); Dr ZL Sun, (szl@5055.cn)

Abstract

The purpose of the current ex vivo study was to compare the speed of vaporesection of human prostatic tissue with benign prostatic hyperplasia (BPH) and the depth of tissue damage using 70- and 120-W 2-μm laser devices. Fresh prostatic tissue specimens were obtained from five patients by open prostatectomy, and were divided into separate groups (70 and 120 W) based on the energy of the laser output (70 and 120 W, respectively). The vaporesection speed, coagulation zone depth and the necrotic tissue layer in the prostatic tissue were evaluated. The current result showed that the speeds (mean±s.d.) of vaporesection were 5.21 ± 0.66 and 10.39 ± 1.15 g/5 min for the 70 and 120 W groups, respectively (P=0.000). There was no difference in the depth of necrosis/coagulation (0.98 ± 0.13/0.30 ± 0.09 and 0.99 ± 0.12/0.31 ± 0.08 mm) for the 70 and 120 W groups, respectively. In conclusion, both 70- and 120-W 2-μm laser devices had superficial tissue damage during the vaporesection of human prostate tissue; moreover, the 120-W laser offers a higher vaporesection speed than the 70-W laser.

Keywords: benign prostatic hyperplasia; vaporesection; 2-μm laser